MSC-1020 Multi Signal Calibrator XXXXXX XXXXXXX | North Protection

MSC-1020 Multi Signal Calibrator XXXXXX XXXXXXX

Description

MSC-1020 is a battery operated instrument that is used to electronically simulate variety of outputs from various types of sensors.

MSC-1020 uses a menu-driven 4×16 character LCD display to establish appropriate settings.

The key panel contains five sealed switches marked with arrows, E (Enter), Back and ON/OFF.

Power is supplied from 4 AA rechargeable, internally mounted Ni-MH batteries which can be recharged with a regulated 9Vdc source.

Connection to PC is established over the front mounted LEMO connector. PC mode will be automatically started after inserting proper cable into device.

Output signal is user selectable from: single-ended voltage (mV), single-ended charge (pC), differential charge (pC), current-sinking IEPE (ICP®), loop powered current source (mA), tachometer (TTL), flow (TTL), OSO® (Optical Speed Output) and BOV (Bias Output Voltage). Frequency range is 1Hz to 10kHz; RMS output voltage is from 10mV to 10V or 10pC to 10.000pC. Outputs can be provided in acceleration, velocity, displacement, voltage or charge.

Application note

Suitable for inspecting measurement lines according to ISO 10816 standards. Device is especially designed for Condition Monitoring Systems (CMS) and/or Vibration Monitoring Systems (VMS) that are independent or connect to SCADA Systems.

 

 

  • Simulates various sensor signals
  • Loop powered current source output
  • Tachometer TTL and OSO® output
  • BOV sensor test
  • Menu-driven operation
  • Multi language menu
  • Metric and imperial units
  • PC connection
  • Battery operated

Outputs

Type

Single-ended Voltage (mV)

IEPE (ICP®) – Current Sinking

Loop powered Current Source (mA)

Single-ended Charge (pC)

Differential Charge (pC)

Tachometer (TTL)

Flow Meter (TTL)

OSO® – Optical Speed Output

BOV – Bias Output Voltage

Frequency range 1Hz to 10kHz

Output range 10mV to 10,000mV RMS; 10pC to 10,000pC

 

Transfer Characteristics

Amplitude accuracy ± 0.5% of settings on any range

Amplitude stability 0.03%/°C maximum change from -10°C to +65°C

Frequency accuracy ± 0.02% of settings on any range

Frequency stability ± 0.5% of maximum change from -10°C to +65°C

Total harmonic distortion ≤ 0.1% 1Hz to 2kHz, ≤ 0.15% 2kHz to 10kHz

 

Environmental Characteristics

Temperature

Operating -10°C to +65°C

Storage -18°C to +65°C

Humidity 95% R.H. maximum

 

Power

Battery 4x AA rechargeable Ni-MH supplied

Autonomy More than 5 hours when fully charged

 

Physical Characteristics

Dimension 196mm x 100mm x 40mm

Weight 0.5kg typical

Case Molded Plastic Case

Connection mV, IEPE, BOV, mA and Tachometer – BNC Charge (single-ended and differential) – Microdot USB – LEMO

Front Panel Controls Five sealed switches (Arrows, E, Back and ON/OFF)

Front Panel Display 4 line LCD panel with 64 character

    NOTE: All technical data can be changed without notice.